LTT SensorCorder is a mobile measurement system with 8 or 16 galvanic isolated Strain Gauge, Volt and ICP inputs that can be set up separately for each sensor type using dedicated software. The inputs are measured with 16 bit resolution.
The maximum sample rate is 500 kHz per channel. The galvanic isolation works in the frequency range from 0 to 50 kHz. Automatic zero balancing, analogue and digital input filtering, internal shunt calibration, quarter, half and full bridge connections and AC/DC coupling round off the system. Ultra-precision constant current and constant voltage supplies are provided for each channel. The internal memory is upgradeable from 4 MS per channel to 16 MS per channel.
The results can be saved directly in common formats like DIAdem, Famos and MGraph, thus avoiding cumbersome export of data. You can start analysing the results immediately upon recording.
The system has been developed especially for ultra-fast long- and short-term measurements. It communicates with a PC via Ethernet (1000/100/10 BaseT), WLAN, SCSI, USB or FireWire. Using the DLL you can integrate other applications (e.g. FAMOS, LabView, DIAdem, MATLAB, Mgraph).
LTT SensorCorder can be customised to your needs. The system is available
- as compact desktop version with 8 or 16 channels
- with cascaded units up to 4096 channels!
- with an optional internal hard drive (40GB) for operation as a stand-alone data recorder
- as 19“ rack mount version in silver design (32 channels per rack), optionally with stand-alone functionality
Application report: LTT's SensorCorder at the gas turbine test stand of Siemens Power Generation
Whitepaper-Download (41 KB)
This product information